Skip to main navigation Skip to search Skip to main content

Image reconstruction with perturbed trajectories in C-arm CT

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

For a 3D C-arm computed tomography systems, actual path of the scanner may deviate from the idea circle geometry because of mechanical instability, leading to perturbation artifacts in reconstructed images. In this paper, we proposed a modified FBP method for the perturbed trajectories taking into account six perturbation parameters without assuming any condition to be ideal. The preliminary studies demonstrated that this algorithm can acquire promising reconstruction image quality even when the perturbations are relatively large. The comparison of performances among different perturbation parameters is useful for constructing a Carm CT system.

Original languageEnglish
Title of host publicationWorld Congress on Medical Physics and Biomedical Engineering
Pages823-826
Number of pages4
DOIs
StatePublished - 2013
EventWorld Congress on Medical Physics and Biomedical Engineering - Beijing, China
Duration: 26 May 201231 May 2012

Publication series

NameIFMBE Proceedings
Volume39 IFMBE
ISSN (Print)1680-0737

Conference

ConferenceWorld Congress on Medical Physics and Biomedical Engineering
Country/TerritoryChina
CityBeijing
Period26/05/1231/05/12

Keywords

  • C-arm CT
  • FBP
  • Trajectory deviation

Fingerprint

Dive into the research topics of 'Image reconstruction with perturbed trajectories in C-arm CT'. Together they form a unique fingerprint.

Cite this