Experimental validation of the accuracy of robotic-assisted radioactive seed implantation for tumor treatment

  • Zhe Ji
  • , Zhe Wang
  • , Yuliang Jiang
  • , Haitao Sun
  • , Bin Qiu
  • , Chuang Li
  • , Bo Liu
  • , Qianqian Li
  • , Leifeng Zhang
  • , Xiangzhi Bai
  • , Ying Wei
  • , Changle Li
  • , Ruoyu Wang*
  • , Junjie Wang*
  • *Corresponding author for this work
  • Peking University
  • Dalian University
  • Beihang University
  • Shandong University
  • Harbin Institute of Technology

Research output: Contribution to journalArticlepeer-review

Abstract

An experimental validation of a robotic system for radioactive iodine-125 seed implantation (RISI) in tumor treatment was conducted using customized phantom models and animal models simulating liver and lung lesions. The robotic system, consisting of planning, navigation, and implantation modules, was employed to implant dummy radioactive seeds into the models. Fiducial markers were used for target localization. In phantom experiments across 40 cases, the mean errors between planned and actual seed positions were 0.98 ± 1.05 mm, 1.14 ± 0.62 mm, and 0.90 ± 1.05 mm in the x, y, and z directions, respectively. The x, y, and z directions correspond to the left–right, anterior–posterior, and superior-inferior anatomical planes. Silicone phantoms exhibiting significantly smaller x-axis errors compared to liver and lung phantoms (p < 0.05). Template assistance significantly reduced errors in all axes (p < 0.05). No significant dosimetric deviations were observed in parameters such as D90, V100, and V150 between plans and post-implant doses (p > 0.05). In animal experiments across 23 liver and lung cases, the mean implantation errors were 1.28 ± 0.77 mm, 1.66 ± 0.69 mm, and 1.86 ± 0.93 mm in the x, y, and z directions, slightly higher than in phantoms (p < 0.05), with no significant differences between liver and lung models. The dosimetric results closely matched planned values, confirming the accuracy of the robotic system for RISI, offering new possibilities in clinical tumor treatment.

Original languageEnglish
Article number219
JournalJournal of Robotic Surgery
Volume18
Issue number1
DOIs
StatePublished - Dec 2024

Keywords

  • Accuracy
  • Experimental validation
  • Radioactive seed implantation
  • Robot
  • Template

Fingerprint

Dive into the research topics of 'Experimental validation of the accuracy of robotic-assisted radioactive seed implantation for tumor treatment'. Together they form a unique fingerprint.

Cite this