Skip to main navigation Skip to search Skip to main content

Error analysis of a computer-assisted orthopedic system for distal locking of intramedullary nail

  • Capital Medical University

Research output: Contribution to journalArticlepeer-review

Abstract

To solve the distal locking difficulty arising from the bending and twisting of the intramedullary nail axis while the nail implanting in long bone fracture therapy, an image-guided computer-assisted orthopaedic surgery system was introduced. To evaluate the error distribution of the system, a special error analysis method was designed based on the character of the system structure and its geometric model. The concrete calculation and statistics of error distribution were embodied by the assistance of the Monte-Carlo method. The system performance was also evaluated in order to derive the theoretical guarantees for further optimizing the system. This error analysis method can be used for the similar systems with complicated mathematical models.

Original languageEnglish
Pages (from-to)850-854
Number of pages5
JournalBeijing Hangkong Hangtian Daxue Xuebao/Journal of Beijing University of Aeronautics and Astronautics
Volume30
Issue number9
StatePublished - Sep 2004

Keywords

  • Computer vision
  • Computer-assisted surgery
  • Intramedullary nail
  • Monte-Carlo method

Fingerprint

Dive into the research topics of 'Error analysis of a computer-assisted orthopedic system for distal locking of intramedullary nail'. Together they form a unique fingerprint.

Cite this