Effects of mechanical load on the degradation of poly(d,l-lactic acid) foam

Research output: Contribution to journalArticlepeer-review

Abstract

Degradable behaviors of polymer for implantation in body should be evaluated before clinical application. The effect of continuous mechanical load on the degradation progress of poly(d,l-lactic acid) (PDLLA) foam gasket was investigated in detail by specially designed load-providing devices. While PDLLA degraded in the PBS solution (pH, 7.4) at 37 °C for 3 months, the changes of surface morphology, molecular weight, elastic modulus, tensile strength and mass loss were recorded. The results revealed that the degradation rates of PDLLA under continuous loads were obviously quicker than those without load. Moreover, the influence of tensile plus compressive load was larger than that of tensile load. It was indicated that in vivo degradation of PDLLA would not only be influenced by the local solution, but also by the surrounding load. When regulating the degradation rate of bioabsorbable polymer, one should consider the indispensable effect of load where implanted.

Original languageEnglish
Pages (from-to)677-683
Number of pages7
JournalPolymer Degradation and Stability
Volume93
Issue number3
DOIs
StatePublished - Mar 2008

Keywords

  • Combined load
  • In vitro degradation
  • PDLLA
  • Tensile load

Fingerprint

Dive into the research topics of 'Effects of mechanical load on the degradation of poly(d,l-lactic acid) foam'. Together they form a unique fingerprint.