Skip to main navigation Skip to search Skip to main content

Analysis of effect of pump light wavelength on the magnetic noise suppression of the SERF co-magnetometer

Research output: Contribution to journalArticlepeer-review

Abstract

Spin Exchange Relaxation-Free Co-Magnetometer (SERFCM) possesses ultra-sensitive rotation measurement capabilities, offering broad application prospects in inertial navigation and fundamental physics research. Precision of SERFCM in rotation measurements is directly influenced by its ability to suppress magnetic noise. In this study, a state-space method is used to establish a SERFCM response model, considering the effect of pump light wavelength. An analysis is conducted to assess the effect of pump light wavelength detuning on the ability to suppress magnetic noise of SERFCM. Theoretical results indicate that detuned pump light wavelength can effectively reduce the magnetic field gradient generated by atomic polarization, thereby improving overall system performance. Experimental validation demonstrates that a slight detuning of the pump light wavelength results in a 21% improvement in the ability to suppress low-frequency magnetic fields.

Original languageEnglish
Article number115471
JournalSensors and Actuators A: Physical
Volume374
DOIs
StatePublished - 16 Aug 2024

Keywords

  • Detuning of the wavelength
  • Magnetic noise
  • Optical pumping
  • Polarization gradient
  • SERF co-magnetometer

Fingerprint

Dive into the research topics of 'Analysis of effect of pump light wavelength on the magnetic noise suppression of the SERF co-magnetometer'. Together they form a unique fingerprint.