Abstract
Heart rate measurement is a basic and important issue for either medical diagnosis or daily health monitoring. In this work great efforts have been focused on realizing a portable, comfortable and low cost solution for long-term domestic heart rate monitoring. A tiny but efficient measurement system composed of a polymer-based flexible pressure sensor and an analog anti-interference readout circuit is proposed; manufactured and tested. The proposed polymer-based pressure sensor has a linear response and high sensitivity of 13.4 kPa-1. With the circuit’s outstanding capability in removing interference caused by body movement and the highly sensitive flexible sensor device, comfortable long-term heart rate monitoring becomes more realistic. Comparative tests prove that the proposed system has equivalent capability (accuracy: <3%) in heart rate measurement to the commercial product.
| Original language | English |
|---|---|
| Pages (from-to) | 3224-3235 |
| Number of pages | 12 |
| Journal | Sensors |
| Volume | 15 |
| Issue number | 2 |
| DOIs | |
| State | Published - 2 Feb 2015 |
| Externally published | Yes |
Keywords
- Anti-interference circuit
- Flexible sensors
- Heart rate monitoring
- Pressure sensors
Fingerprint
Dive into the research topics of 'A pressure sensing system for heart rate monitoring with polymer-based pressure sensors and an anti-interference post processing circuit'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver