A practical method to calibrate the slant angle of central X-ray for laminography scanning system

  • Min Yang*
  • , Jianhua Zhu
  • , Qi Liu
  • , Shengling Duan
  • , Lihong Liang
  • , Xingdong Li
  • , Wenli Liu
  • , Fanyong Meng
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

Computed laminography (CL) has a particular function in plate-type structure tomography. For CL reconstruction, slant angle of the central X-ray is an important parameter and must be calibrated accurately. A practical feasible calibration method to determine the slant angle is proposed. According to this method, a spherical phantom is fixed on the rotary table and turns a full 360 rotation. Then, image processing and the least-square fitting are performed to figure out the locus of the projection. Consequently, the slant angle is achieved in accordance with the property of the fitted ellipse. The experimental results prove that this method is easy to implement and meet the inspection requirements.

Original languageEnglish
Pages (from-to)13-20
Number of pages8
JournalNDT and E International
Volume64
DOIs
StatePublished - Jun 2014

Keywords

  • Computed laminography (CL)
  • Ellipse fitting
  • Geometrical calibration
  • Slant angle