A dual‐scintillator time‐of‐flight spectrometer for secondary fast neutrons in proton radiotherapy

  • Wei Cheng
  • , Wenbao Jia
  • , Daqian Hei
  • , Weiwei Qu*
  • , Lian Chen
  • , Gaolong Zhang
  • *Corresponding author for this work
  • Nanjing University of Aeronautics and Astronautics
  • Collaborative Innovation Center of Radiation Medicine of Jiangsu Higher Education Institutions
  • Lanzhou University
  • Soochow University
  • University of Science and Technology of China

Research output: Contribution to journal › Article › peer-review

Abstract

In order to measure the energy spectra of secondary neutrons in proton and heavy-ion radiotherapy, a neutron spectrometer has been developed. In this paper, two organic scintillation detectors combined with photomultiplier tubes were used to form a dual-scintillator time-of-flight spectrometer. The performance of this spectrometer was tested with a DT neutron generator, and the detection efficiency of each detector as well as the spectrometer were simulated by Geant4 toolkit. It’s shown that the time resolution of the spectrometer is 0.797 ns with an energy resolution of 26.4% for neutrons with an energy of 14 MeV at a flight distances over 150 cm.

Original languageEnglish
Pages (from-to)1317-1323
Number of pages7
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume327
Issue number3
DOIs
StatePublished - Mar 2021

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Neutron energy
  • Pulse shape discrimination
  • Scintillation detector
  • Time‐of‐flight