Abstract
The standardization of grayscale display is essentially significant for image signal communication, transmission, and terminal reading. The key step of this standardization is establishing a traceable equipment of grayscale. As a relative value, grayscale is transferred to two different absolute values to satisfy different traceability methods, including optical density for hardcopy image and luminance for softcopy. For luminance, a generation equipment is designed to build the relationship between luminance and grayscale. In this work, novel equipment is established using digital light processing (DLP) by time-frequency modulation, and the corresponding uncertainty is analyzed. The experiment result shows that this digital equipment builds the relationship between grayscale and luminance in the range of 0.16-4000 cd/m2. It enables traceable measurement of grayscale to luminance on this equipment with high accuracy and can provide a standardized reference for the display of grayscale images in the fields of medicine, remote sensing, non-destructive testing, etc.
| Original language | English |
|---|---|
| Article number | 2297 |
| Journal | Applied Sciences (Switzerland) |
| Volume | 10 |
| Issue number | 7 |
| DOIs | |
| State | Published - 1 Apr 2020 |